Skip to main content

Table 5 Associations of poor tolerability with removal of monitor and inadequate data acquisition

From: Tolerability of the Oscar 2 ambulatory blood pressure monitor among research participants: a cross-sectional repeated measures study

  Poor tolerability to wearing monitor
  Yes (%) No (%) p-value
Day removal, first wearing 7.7 2.8 0.44
Night removal, first wearing 15.4 2.9 0.11
Day removal, second wearing 14.3 5.0 0.25
Night removal, second wearing 28.6 2.5 0.004
Inadequate data for daytime, first wearing 15.4 2.8 0.10
Inadequate data for nighttime, first wearing 23.1 19.4 0.78
Inadequate data for entire 24 hours, first wearing 23.1 19.4 0.78
Inadequate data for daytime, second wearing 21.4 2.5 0.02
Inadequate data for nighttime, second wearing 42.9 22.5 0.14
Inadequate data for entire 24 hours, second wearing 50.0 22.5 0.05